DOI 10.1093/bja/aes142

Медицинская библиотека · British journal of anaesthesia · 2012

Cadaver study of oesophageal insufflation with supraglottic airway devices during positive pressure ventilation in an obstructed airway.

Schmidbauer W, Genzwürker H, Ahlers O, Proquitte H, Kerner T

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Supraglottic airway devices (SADs) play an increasing role in airway management in clinical anaesthesia and emergency medicine. Until now, no data exist concerning the extent of oesophageal insufflation when oropharyngeal leak pressures are exceeded.

METHODS

Laryngeal masks LMA-Supreme™ and LMA-ProSeal™, laryngeal tubes LTS-D and LTS II, Combitube™, and I-Gel were inserted into unfixed human cadavers. The oesophagus was connected to a volumeter, while the trachea was closed surgically to simulate complete airway obstruction. Volumes of oesophageal insufflation resulting from pressure-controlled ventilation at inspiratory pressures of 20, 40, and 60 mbar were measured.

RESULTS

No oesophageal insufflation could be detected at a ventilation pressure of 20 mbar in any device. Using inspiratory pressures of 40 and 60 mbar, oesophageal insufflation occurred in all devices, with significantly higher volumes of intraoesophageal air for both laryngeal tubes.

CONCLUSIONS

The use of SADs with inspiratory pressures of 20 mbar appears to be safe regarding the risk of intragastric insufflation. Higher inspiratory pressures should be strictly avoided.

Для цитирования: Schmidbauer W, Genzwürker H, Ahlers O, Proquitte H, Kerner T. Cadaver study of oesophageal insufflation with supraglottic airway devices during positive pressure ventilation in an obstructed airway.. British journal of anaesthesia. 2012. DOI: 10.1093/bja/aes142.

Предметные рубрики MeSH