Cornea · 2024
A Novel Artificial Endothelial Replacement Membrane for the Treatment of Chronic Corneal Edema.
Клиническое исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Daphna Ofer, Auffarth Gerd U, Lapid-Gortzak Ruth, Chaurasia Sunita, Gilboa Efrat, Lemze Anat, Dover Michael, Marcovich Arie L
Аннотация
PURPOSE: The purpose of this study was to report the safety and efficacy results of an artificial lamellar implant for the treatment of chronic corneal edema.
METHODS: The EndoArt (EyeYon Medical, Ness Ziona, Israel), an artificial endothelial replacement membrane designed to treat corneal edema, was implanted in 24 eyes of 24 patients with low-to-normal visual potential. We present the safety and efficacy results from a prospective, open-label, single-arm, multicenter study conducted over a 12-month period.
RESULTS: Twenty-four patients were enrolled, with no device-related serious adverse events reported. Seventeen patients completed 12-month follow-up, showing a reduction in average central corneal thickness from 759 ± 116 μm to 613 ± 135 μm. Best-corrected distance visual acuity improved from 1.88 ± 0.79 logarithmic minimum angle of resolution (logMAR) to 1.34 ± 0.57 logMAR. Sixty percent gained at least 3 early treatment diabetic retinopathy study (ETDRS) lines. The EndoArt was removed in 5 cases due to incomplete attachment and replaced by corneal transplants; 1 patient was lost to follow-up, and 1 had a procedure failure. No device-related long-term complications, infections, or inflammations were reported. The implants remained transparent throughout the study.
CONCLUSIONS: The first-in-human results of EndoArt implantation demonstrated the device's potential to treat patients suffering from corneal edema with a favorable safety profile and effective edema reduction in most subjects, with no device-related serious adverse event. The EndoArt may offer a viable solution in regions facing a shortage of donor corneas, as well as for patients who have poor prognosis with human tissue.
METHODS: The EndoArt (EyeYon Medical, Ness Ziona, Israel), an artificial endothelial replacement membrane designed to treat corneal edema, was implanted in 24 eyes of 24 patients with low-to-normal visual potential. We present the safety and efficacy results from a prospective, open-label, single-arm, multicenter study conducted over a 12-month period.
RESULTS: Twenty-four patients were enrolled, with no device-related serious adverse events reported. Seventeen patients completed 12-month follow-up, showing a reduction in average central corneal thickness from 759 ± 116 μm to 613 ± 135 μm. Best-corrected distance visual acuity improved from 1.88 ± 0.79 logarithmic minimum angle of resolution (logMAR) to 1.34 ± 0.57 logMAR. Sixty percent gained at least 3 early treatment diabetic retinopathy study (ETDRS) lines. The EndoArt was removed in 5 cases due to incomplete attachment and replaced by corneal transplants; 1 patient was lost to follow-up, and 1 had a procedure failure. No device-related long-term complications, infections, or inflammations were reported. The implants remained transparent throughout the study.
CONCLUSIONS: The first-in-human results of EndoArt implantation demonstrated the device's potential to treat patients suffering from corneal edema with a favorable safety profile and effective edema reduction in most subjects, with no device-related serious adverse event. The EndoArt may offer a viable solution in regions facing a shortage of donor corneas, as well as for patients who have poor prognosis with human tissue.