Real-time imaging of transscleral diode laser cyclophotocoagulation by optical coherence tomography.
Аннотация
PURPOSE: To present optical coherence tomography (OCT) for real-time imaging of cyclophotocoagulation effects.
METHODS
In a pilot study, real-time transscleral OCT images were generated during diode laser cyclophotocoagulation in four eyes of four patients suffering from uncontrolled glaucoma using a specially designed contact applicator containing the OCT fiber, a focussing fiber optic and the fiber of the diode laser.
RESULTS
When the contact system was used, two layers could be differentiated: a superficial thick hyperreflective complex representing conjunctiva, Tenon's capsule, episclera and sclera, and a thinner hyporeflective layer representing the ciliary body. During cyclophotocoagulation, real-time OCT showed a clear and sudden thickening of the ciliary body in the treated area.
CONCLUSION
This new OCT device represents a first step towards visual, real-time imaging of cyclophotocoagulation. After further adaptation of the delivery system, further trials are needed to correlate OCT findings with aqueous production and intraocular pressure.