DOI 10.3233/thc-231338

Медицинская библиотека · Technology and health care : official journal of the European Society for Engineering and Medicine · 2024

Portable technology for postural control measurement: Comparing head position with center of pressure data.

Harel Daphna, Lubetzky Anat Vilnai

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Standing is a basic human function that healthy adults take for granted, yet it is a complex perceptual-motor process that requires sensation of position and motion from the sensory systems.

OBJECTIVE

We assessed agreement between center of pressure data from a laboratory force-platform and head position data from an HTC Vive head-mounted display (HMD) for the evaluation of standing postural control. We investigated the impact of different statistical choices when assessing the relationship between two measurements. Specifically: 1) How does correlation and agreement statistics relate before and after logarithmic transformation? 2) Is there systemic or proportional bias between the force-platform and HMD measurements?

METHODS

We tested 37 adults (26 controls, 11 with unilateral vestibular hypofunction) standing on foam, observing a static or dynamic visual scene projected from the HMD. We quantified anterior-posterior and medio-lateral sway via Directional Path, Root Mean Square Velocity, Variance, and Power Spectral Density (PSD) from a force-platform and the HMD.

RESULTS

Intra-class correlations (ICCs) were moderate-to-good for the non-transformed data and good-to-excellent after logarithmic transformation for all outcomes except for PSD above 1 Hz. Correlations were higher than ICCs. Bland-Altman plots indicated proportional bias but not after logarithmic transformation.

CONCLUSIONS

Both devices correlated linearly, and measure people's postural responses but cannot be used interchangeably, mostly because they appear to diverge with larger sway as evident on Bland-Altman plots of non-transformed data. Agreement between devices was excellent for low frequency movement but poor for high frequency small corrective movements.

Для цитирования: Harel Daphna, Lubetzky Anat Vilnai. Portable technology for postural control measurement: Comparing head position with center of pressure data.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2024. DOI: 10.3233/thc-231338.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.