High-Sensitivity Ferrite Magnetometer for Low-Dose Magnetic Sentinel Lymph Node Detection.

Wang Qibin, Chen Xinyue, Li Lei, Yang Kaixiong, Hou Lingwen, Shu Wang, Wu Zhongyi, Feng Xin, Zhu Shouping

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Magnetic tracer-guided sentinel lymph node (SLN) biopsy offers a radiation-free alternative for nodal staging in breast cancer and other solid tumors. However, current handheld magnetometers often require high superparamagnetic iron oxide nanoparticle (SPION) doses due to limited sensitivity and detection depth, leading to skin staining and MRI artifacts.

METHODS

We developed FerriMag, a handheld ferrite magnetometer integrating a central high-permeability Mn-Zn ferrite core to enhance magnetic excitation and signal reception. Lock-in amplification and bandpass filtering were employed to suppress interference.

RESULTS

The ferrite core increased the excitation field strength by 11-fold and receiving-coil sensitivity by 1.7-fold. FerriMag achieved a detection limit of 21.6 ng Fe and reliably detected 0.5 μg Fe at a depth of 10 mm. In a rabbit model, accurate SLN localization was obtained with only 0.05 mL (0.5 mg Fe) of SPIONs, without visible skin staining. Metastatic nodes exhibited significantly reduced SPION uptake (p = 0.0005) and slower signal rise(p = 0.038) compared to non-metastatic nodes.

CONCLUSION

FerriMag improves the sensitivity and detection depth of handheld magnetic SLN detection, enabling reliable low dose SPION-based SLN localization in vivo.

SIGNIFICANCE: This system provides a low-cost, radiation-free, and clinically translatable platform for low-dose magnetic SLN detection, with potential applicability beyond breast cancer and possible utility for intraoperative nodal assessment.

Для цитирования: Wang Qibin, Chen Xinyue, Li Lei, Yang Kaixiong, Hou Lingwen, Shu Wang, Wu Zhongyi, Feng Xin, Zhu Shouping. High-Sensitivity Ferrite Magnetometer for Low-Dose Magnetic Sentinel Lymph Node Detection.. IEEE transactions on bio-medical engineering. 2026. DOI: 10.1109/tbme.2026.3719508.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.