Pulmonary Cryoablation Zones: More Aggressive Ablation Is Warranted In Vivo.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lyons Gray R, Winokur Ronald S, Pua Bradley B

Аннотация

OBJECTIVE: The objective of our study was to determine the effective cryoablation zone when treating pulmonary tumors in vivo and to create pulmonary-specific ablation maps to guide clinical procedure planning.

MATERIALS AND METHODS: Ablation volume was measured retrospectively in human patients after pulmonary tumor cryoablation with a triple-freeze protocol. Single-probe ablations were performed with 17-, 14-, and 13-gauge cryoprobes; multiple-probe ablations were performed with two or three 17-gauge probes. Statistical comparisons of ablation volumes to manufacturer reference values were calculated using the Wilcoxon rank-sum test. Comparisons of ablation sizes by the number of probes were evaluated by the Kruskal-Wallis test.

RESULTS: Mean volume of in vivo lung ablation with a single 17-gauge cryoprobe measured 3.0 cm3, which is a statistically significant difference compared with the in vitro -20°C isotherm volume of 22.6 cm3 (p < 0.01). Mean ablation volume of larger 13- and 14-gauge cryoprobes were 4.3 and 1.8 cm3, respectively, both of which are smaller than the in vitro -20°C isotherm volume. Mean cryoablation zone was not significantly affected by distance to the pleura (p = 0.54) or distance to a vessel (p = 0.55). Ablation volume was significantly increased (p < 0.01) with the use of multiple cryoprobes, at a rate of a 10.8-cm3 increase per additional probe. The increased ablation zone size was more attributable to increased short-axis width (9.6-mm increase per probe) compared with long-axis length (5.6-mm increase per probe).

CONCLUSION: The in vivo effective pulmonary cryoablation zone is significantly smaller than the manufacturer-published in vitro isotherm. Larger ablation margins in lung are best achieved by using multiple cryoprobes.