Pulmonary Cryoablation Zones: More Aggressive Ablation Is Warranted In Vivo.
Аннотация
OBJECTIVE
The objective of our study was to determine the effective cryoablation zone when treating pulmonary tumors in vivo and to create pulmonary-specific ablation maps to guide clinical procedure planning.
MATERIALS AND
METHODS
Ablation volume was measured retrospectively in human patients after pulmonary tumor cryoablation with a triple-freeze protocol. Single-probe ablations were performed with 17-, 14-, and 13-gauge cryoprobes; multiple-probe ablations were performed with two or three 17-gauge probes. Statistical comparisons of ablation volumes to manufacturer reference values were calculated using the Wilcoxon rank-sum test. Comparisons of ablation sizes by the number of probes were evaluated by the Kruskal-Wallis test.
RESULTS
Mean volume of in vivo lung ablation with a single 17-gauge cryoprobe measured 3.0 cm3, which is a statistically significant difference compared with the in vitro -20°C isotherm volume of 22.6 cm3 (p < 0.01). Mean ablation volume of larger 13- and 14-gauge cryoprobes were 4.3 and 1.8 cm3, respectively, both of which are smaller than the in vitro -20°C isotherm volume. Mean cryoablation zone was not significantly affected by distance to the pleura (p = 0.54) or distance to a vessel (p = 0.55). Ablation volume was significantly increased (p < 0.01) with the use of multiple cryoprobes, at a rate of a 10.8-cm3 increase per additional probe. The increased ablation zone size was more attributable to increased short-axis width (9.6-mm increase per probe) compared with long-axis length (5.6-mm increase per probe).
CONCLUSION
The in vivo effective pulmonary cryoablation zone is significantly smaller than the manufacturer-published in vitro isotherm. Larger ablation margins in lung are best achieved by using multiple cryoprobes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.