m+Medelevator.RU
← Поиск публикаций
Journal of refractive surgery (Thorofare, N.J. : 1995) · 2005

Potentially accommodating intraocular lenses--an in vitro and in vivo study using three-dimensional high-frequency ultrasound.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Stachs Oliver, Schneider Hanka, Stave Joachim, Guthoff Rudolf

Аннотация

PURPOSE: To investigate the accommodative performance of new intraocular lenses (IOL) using the advantages of three-dimensional ultrasound biomicroscopy.

METHODS: An in vitro simulation device was designed to study IOL performance using an artificial capsular bag and a stretching device. The haptic region of the Akkommodative 1CU (HumanOptics AG) and CrystaLens AT-45 (Eyeonics Inc) was visualized in vitro in three dimensions, using an in-house developed three-dimensional ultrasound biomicroscope. The in vitro results were used to describe the in vivo situation in four patients with accommodative implants.

RESULTS: The haptic position and angulation in consideration of the accommodation state was distinguished and analyzed. In the simulation model, a maximal angulation change of 4.5 degrees and 4.3 degrees and a maximal forward shift of 0.33 mm and 0.28 mm was observed for the AT-45 and 1CU, respectively. In vivo, a change in haptic angulation <100 and a maximal forward shift of 0.50 mm was observed for the 1CU. These changes correspond to a theoretical approximate value of 0.50 diopters.

CONCLUSIONS: The in vitro simulation device examined with three-dimensional ultrasound biomicroscopy provided information on the accommodative performance of these potentially accommodative IOL designs. Using three-dimensional ultrasound biomicroscopy, corresponding changes in haptic angulation during pharmacological-induced accommodation were observed.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты