MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/aln.0b013e31819c45e5

Медицинская библиотека · Anesthesiology · 2009

Use of the Pentax-AWS in 293 patients with difficult airways.

Asai Takashi, Liu Eugene H, Matsumoto Sanae, Hirabayashi Yoshihiro, Seo Norimasa, Suzuki Akihiro, Toi Takashi, Yasumoto Kazumasa, Okuda Yasuhisa

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Several case reports have shown that the Pentax-AWS (Hoya Corporation, Tokyo, Japan), a new video laryngoscope, is useful in patients with difficult airways.

METHODS

We assessed the effectiveness of the Pentax-AWS in two groups. Group 1 included 270 patients in whom direct laryngoscopy using a Macintosh laryngoscope had been difficult. Group 2 included 23 patients with predicted difficult intubation and difficult mask ventilation without previous use of the Macintosh laryngoscope.

RESULTS

In group 1, the view of the glottis with the Macintosh laryngoscope was Cormack and Lehane grade 2 in 14 patients, grade 3 in 208 patients, and grade 4 in 48 patients. In 256 patients in whom the grade was 3 or 4 with the Macintosh laryngoscope, the view with the Pentax-AWS was either grade 1 or 2 in 255 patients (99.6%; 95% confidence intervals 97.8-100%). Tracheal intubation was successful with the Pentax-AWS in 268 of 270 patients (99.3%; 95% confidence interval 97.4-100%), and it failed (after two attempts) in two patients. In group 2, tracheal intubation was successful in 22 of 23 patients, and it failed in one patient. The reasons for failed intubation using the Pentax-AWS were failure to position the blade toward the glottic side of the epiglottis, inability to maneuver the endotracheal tube away from the arytenoids and into the trachea, and bleeding and swelling of the oropharynx.

CONCLUSION

The success rate of tracheal intubation using the Pentax-AWS was high in patients with difficult laryngoscopy with a Macintosh laryngoscope and in patients with predicted difficult intubation.

Для цитирования: Asai Takashi, Liu Eugene H, Matsumoto Sanae, Hirabayashi Yoshihiro, Seo Norimasa, Suzuki Akihiro, Toi Takashi, Yasumoto Kazumasa, Okuda Yasuhisa. Use of the Pentax-AWS in 293 patients with difficult airways.. Anesthesiology. 2009. DOI: 10.1097/aln.0b013e31819c45e5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.