Configuration analysis of the injection position and shape of the gel spacer in gynecologic brachytherapy.

Iijima Kotaro, Murakami Naoya, Nakamura Satoshi, Nishioka Shie, Chiba Takahito, Kuwahara Junichi, Takemori Mihiro, Nakayama Hiroki, Kashihara Tairo, Okuma Kae, Takahashi Kana, Inaba Koji, Igaki Hiroshi, Okamoto Hiroyuki, Itami Jun

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: In this single-institution retrospective study, configuration analysis was performed to determine the optimal location and volume of hyaluronic acid gel spacer injection into the rectovaginal or vesicovaginal septum for effective dose reduction (DR) to the organs at risk (OARs), the rectum and bladder.

METHODS AND MATERIALS: 70 and 50 intracavitary brachytherapy treatment plans used only vaginal cylinders with gel spacers for the rectal and bladder sides, respectively, whereas 28 did not use spacers. Correlation analysis was performed between the geometrical parameters and injection position of the gel spacers and the 2-cm3 covering doses of the OARs for each treatment.

RESULTS

A higher DR was predicted for hyaluronic acid gel spacer injection within ±5 mm and ±2.5 mm in the lateral-medial direction from the midpoint on the rectal and bladder sides, and ±10 mm in the cranial-caudal direction from the midpoint on the rectal side. There were correlations between 2-cm3 covering doses and the gel spacer parameters: the volume on the rectal (p = 0.02) and bladder (p = 0.04) sides; the craniocaudal length on the rectal side (p << 0.05); and ventrodorsad thickness on each OAR (p << 0.05) sides. There was no significant difference in the DR between a volume of ∼10 cm3 and that of a higher volume (p >> 0.05).

CONCLUSIONS

A gel spacer volume of ∼10 cm3 provides sufficient OAR DR if its gravity point is on the midpoint between the cylinder applicator and OAR, and its craniocaudal length covers the active length of the cylinder applicator.

Для цитирования: Iijima Kotaro, Murakami Naoya, Nakamura Satoshi, Nishioka Shie, Chiba Takahito, Kuwahara Junichi, Takemori Mihiro, Nakayama Hiroki, Kashihara Tairo, Okuma Kae, Takahashi Kana, Inaba Koji, Igaki Hiroshi, Okamoto Hiroyuki, Itami Jun. Configuration analysis of the injection position and shape of the gel spacer in gynecologic brachytherapy.. Brachytherapy. 2020. DOI: 10.1016/j.brachy.2020.08.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.