MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/01.ju.0000181363.41598.47

Медицинская библиотека · The Journal of urology · 2005

Kidney displacement simulator for retroperitoneal laparoscopic nephrectomy.

Takiuchi Hidekazu, Mori Yoshinori, Shima Hiroki, Tanooka Masao, Hirayama Shin-Ichi, Nakao Norio

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: We evaluated the efficacy of a renal displacement simulator originally developed at our department for retroperitoneal laparoscopic nephrectomy.

MATERIALS AND

METHODS

A total of 12 patients with a malignant localized renal (7) or ureteral (5) neoplasm underwent multidetector row computerized tomography. Imaging data were sent to a dedicated work station to create volume rendering and virtual laparoscopic images of the kidney, which was displaced ventral using a retroperitoneal balloon. These findings were compared with video images obtained during laparoscopy surgery.

RESULTS

The kidney displacement simulator depicted all renal arteries (100% sensitivity) and 13 of 14 renal veins (93% sensitivity). Hilar anatomy, including the tumor, as well as major vessels and their relationships were visualized by the simulator in the laparoscopic views. The major vessel portions completely corresponded to those seen during surgery, and the left adrenal and gonadal veins were also synchronized quite well.

CONCLUSIONS

Our kidney displacement simulator was able to visualize the major vessel portions and branched small vessels, such as the adrenal and gonadal veins, prior to surgery. It is considered useful for providing guidance to surgeons and decreasing operative risks and possible complications.

Для цитирования: Takiuchi Hidekazu, Mori Yoshinori, Shima Hiroki, Tanooka Masao, Hirayama Shin-Ichi, Nakao Norio. Kidney displacement simulator for retroperitoneal laparoscopic nephrectomy.. The Journal of urology. 2005. DOI: 10.1097/01.ju.0000181363.41598.47.