BiClamp forceps significantly shorten the operation time for breast surgery.
Аннотация
PURPOSE: A novel approach was introduced for breast surgery using the BiClamp, a new bipolar thermal energy device, to avoid complications and to shorten the time required for the dissection of the axillary lymph nodes.
METHODS
Thirty-six patients with early breast cancer were assessed. The surgical parameters were compared between the procedures performed using the BiClamp technique (n = 14) and conventional surgery with suture ligation (n = 22). The parameters included the operation time, blood loss, and discharge on the first postoperative day. In addition, each of those parameters was compared between the patients with a high body mass index (BMI) (>22) and a low BMI (< or =22). The sealed vessels were examined histologically and heat-associated morphological vessel wall alterations were evaluated.
RESULTS
The operation time was significantly shorter in the BiClamp group than in the control group (P = 0.017, 90 +/- 18 vs 115 +/- 33 min). In addition, the blood loss in the BiClamp group tended to be smaller than in the control group, but the difference was not significant (P = 0.54, 61 +/- 47 vs 74 +/- 67 g). No other parameters showed any significant differences between the two groups.
CONCLUSION
The BiClamp thermofusion technique was safe and useful in breast surgery involving axillary dissection.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.