High-performance Liquid Chromatographic Ultraviolet Detection of Nilotinib in Human Plasma from Patients with Chronic Myelogenous Leukemia, and Comparison with Liquid Chromatography-Tandem Mass Spectrometry.

Nakahara Ryosuke, Satho Yuhki, Itoh Hiroki

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

A method for determining nilotinib concentration in human plasma is proposed using high-performance liquid chromatography and ultraviolet detection.

MATERIALS &

METHODS

Nilotinib and the internal standard dasatinib were separated using a mobile phase of 0.5% Na2 PO4 H2 O (pH 2.5)-acetonitrile-methanol (55:25:20, v/v/v) on a Capcell Pak C18 MG II column (250 × 4.6 mm) at a flow rate of 1.0 ml/min, and ultraviolet measurement at 250 nm.

RESULTS

The calibration curve exhibited linearity over the nilotinib concentration range of 50-2,500 ng/ml at 250 nm, with relative standard deviations (n = 5) of 7.1%, 2.5%, and 2.9% for 250, 1,500, and 2,500 ng/ml, respectively. The detection limit for nilotinib was 5 ng/ml due to three blank determinations (ρ = 3).

CONCLUSION

This method was successfully applied to assaying nilotinib in human plasma samples from patients with chronic myelogenous leukemia. In addition, we compared the results with those measured by liquid chromatography-tandem mass spectrometry (LC-MS/MS) at BML, Inc. (a commercial laboratory). A strong correlation was observed between the nilotinib concentrations measured by our high-performance liquid chromatographic method and those obtained by LC/MS-MS (r2 = 0.988, P < 0.01).

Для цитирования: Nakahara Ryosuke, Satho Yuhki, Itoh Hiroki. High-performance Liquid Chromatographic Ultraviolet Detection of Nilotinib in Human Plasma from Patients with Chronic Myelogenous Leukemia, and Comparison with Liquid Chromatography-Tandem Mass Spectrometry.. Journal of clinical laboratory analysis. 2016. DOI: 10.1002/jcla.21975.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.