Commercial Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Molecular Assays: Superior Analytical Sensitivity of cobas SARS-CoV-2 Relative to NxTAG CoV Extended Panel and ID NOW COVID-19 Test.

Jin Run, Pettengill Matthew A, Hartnett Nicole L, Auerbach Herbert E, Peiper Stephen C, Wang Zixuan

СтатьяПеревод пока не подготовлен

Аннотация

CONTEXT.—: We implemented multiple nucleic acid amplification test platforms because of the limited availability of test kits for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) during the early stages of the pandemic. Interpretation of results generated by different platforms and prioritization for testing algorithms required cross-comparison.

OBJECTIVE.—: To compare the analytical sensitivity of 3 commercial SARS-CoV-2 molecular assays, selected samples were studied in parallel with Cobas SARS-CoV-2 test, NxTAG CoV Extended Panel, and ID NOW COVID-19 assays.

DESIGN.—: A total of 8043 SARS-CoV-2 tests performed from March 22 to April 19, 2020, were included in this study. For all 1794 positive specimens detected by the cobas SARS-CoV-2 assay, the cycle threshold (Ct) values were manually tracked and plotted to demonstrate the distribution of sample viral levels. Additionally, 50 and 63 low-positive specimens (Ct values >32) as well as 50 and 61 consecutive positive specimens by the cobas assay were tested with NxTAG and ID NOW, respectively, to estimate their relative sensitivities.

RESULTS.—: The Ct values of cobas SARS-CoV-2-positive samples were evenly distributed throughout ranges of 13.32 to 39.50 (mean, 25.06) and 13.60 to 42.49 (mean, 26.45) for ORF1 and E gene targets, respectively. NxTAG reliably detected only specimens with E gene Ct values lower than 33, and is estimated to detect 89.4% of positive specimens detected by cobas assay. ID NOW had performance variation independent of Ct value and is estimated to detect 83.5% of cobas positives.

CONCLUSIONS.—: Clinical specimens exhibit a wide range of viral burden, with a significant portion at low levels. Analytical sensitivity of testing platforms is critical for reliable detection of SARS-CoV-2 and uniform care to patients.

Для цитирования: Jin Run, Pettengill Matthew A, Hartnett Nicole L, Auerbach Herbert E, Peiper Stephen C, Wang Zixuan. Commercial Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Molecular Assays: Superior Analytical Sensitivity of cobas SARS-CoV-2 Relative to NxTAG CoV Extended Panel and ID NOW COVID-19 Test.. Archives of pathology & laboratory medicine. 2020. DOI: 10.5858/arpa.2020-0283-sa.