DOI 10.1515/cclm-2020-0749

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2020

SARS-CoV-2 RNA identification in nasopharyngeal swabs: issues in pre-analytics.

Basso Daniela, Aita Ada, Navaglia Filippo, Franchin Elisa, Fioretto Paola, Moz Stefania, Bozzato Dania, Zambon Carlo-Federico, Martin Barbara, Dal Prà Chiara, Crisanti Andrea, Plebani Mario

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The direct identification of SARS-CoV-2 RNA in nasopharyngeal swabs is recommended for diagnosing the novel COVID-19 disease. Pre-analytical determinants, such as sampling procedures, time and temperature storage conditions, might impact on the end result. Our aim was to evaluate the effects of sampling procedures, time and temperature of the primary nasopharyngeal swabs storage on real-time reverse-transcription polymerase chain reaction (rRT-PCR) results.

METHODS

Each nasopharyngeal swab obtained from 10 hospitalized patients for COVID-19 was subdivided in 15 aliquots: five were kept at room temperature; five were refrigerated (+4 °C); five were immediately mixed with the extraction buffer and refrigerated at +4 °C. Every day and for 5 days, one aliquot per condition was analyzed (rRT-PCR) for SARS-CoV-2 gene E and RNaseP and threshold cycles (Ct) compared. To evaluate manual sampling, 70 nasopharyngeal swabs were sampled twice by two different operators and analyzed separately one from the other.

RESULTS

A total of 6/10 swabs were SARS-CoV-2 positive. No significant time or storage-dependent variations were observed in SARS-CoV-2 Ct. Re-sampling of swabs with SARS-CoV-2 Ct lower than 33 resulted in highly reproducible results (CV=2.9%), while a high variability was observed when Ct values were higher than 33 (CV=10.3%).

CONCLUSIONS

This study demonstrates that time and temperature of nasopharyngeal swabs storage do not significantly impact on results reproducibility. However, swabs sampling is a critical step, and especially in case of low viral load, might be a potential source of diagnostic errors.

Для цитирования: Basso Daniela, Aita Ada, Navaglia Filippo, Franchin Elisa, Fioretto Paola, Moz Stefania, Bozzato Dania, Zambon Carlo-Federico, Martin Barbara, Dal Prà Chiara, Crisanti Andrea, Plebani Mario. SARS-CoV-2 RNA identification in nasopharyngeal swabs: issues in pre-analytics.. Clinical chemistry and laboratory medicine. 2020. DOI: 10.1515/cclm-2020-0749.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.