DOI 10.1007/s10096-020-03905-9

Медицинская библиотека · European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology · 2020

Improved performances of the second generation of the ID NOW influenza A&B 2® and comparison with the GeneXpert®.

Farfour Eric, Roux Antoine, Ballester Marie, Gagneur Léa, Renaux Claudine, Jolly Emilie, Vasse Marc

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Due to the congestion of emergency wards, a rapid and accurate method for flu diagnosis is required. However, the first evaluations of the ID NOW® influenza A&B performances were heterogeneous. We aimed to (i) assess the performance of the second generation of the ID NOW® influenza A&B 2 and (ii) compare the ID NOW® to the GeneXpert® use when performed within a central laboratory. (i) Analytical performances of ID NOW® were assessed using a collection of 112 clinical samples in comparison with a reference multiplex PCR (Seegene®); (ii) Invalid rate per operator and time to result were calculated for the GeneXpert® Xpress Flu/RSV and the ID NOW® influenza A&B 2® during 2017-2018 and 2018-2019 flu outbreaks respectively. ID NOW® reaches an overall sensitivity and specificity of 96.6% and 96.1% respectively. Most of the false-negative display a CT > 37. For both instruments, a single operator was involved in about a half of all invalid results. Excluding this operator involved in most invalid result, the invalid rate was about 1% for both instruments. Time to result from sampling was significantly shorter in the ID NOW® versus the GeneXpert® group (33 vs 97 min, P < 0.01). The second generation of the ID NOW® influenza A&B 2 displays high performances, comparable with conventional PCR method. In order to prevent invalid results, we highlight the need for adequate training of operators. Also, when implemented in a central laboratory, the location of the instrument could have a strong impact on the time-to-results.

Для цитирования: Farfour Eric, Roux Antoine, Ballester Marie, Gagneur Léa, Renaux Claudine, Jolly Emilie, Vasse Marc. Improved performances of the second generation of the ID NOW influenza A&B 2® and comparison with the GeneXpert®.. European journal of clinical microbiology & infectious diseases : official publication of the European Society of Clinical Microbiology. 2020. DOI: 10.1007/s10096-020-03905-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.