DOI 10.1097/lbr.0000000000000444
Cryobiopsy of the Pleura: An Improved Diagnostic Tool.
Аннотация
BACKGROUND
Medical thoracoscopy (rigid and semirigid pleuroscopy) has revolutionized the approach to the diagnosis of pleural disease by offering a very high diagnostic yield. Rigid pleuroscopy offers the advantages of therapeutic intervention and larger biopsy specimens, whereas semirigid pleuroscopy using a standard biopsy forceps yields smaller and more superficial pleural samples. Cryobiopsy through semirigid pleuroscope in anecdotal studies has been used to overcome these disadvantages. We compared the safety and efficacy of cryobiopsy with conventional forceps biopsy in terms of the specimen size and diagnostic yield.
METHODS
We analyzed data of 139 (87 cryobiopsies and 52 forceps biopsies) patients with undiagnosed pleural effusion who underwent pleuroscopy using a semirigid pleuroscope. A cryoprobe (ERBE, 2.4 mm) was passed through the working channel of the semirigid pleuroscope, the target area of parietal pleura was frozen for an average freezing time of 8 seconds, then the semirigid pleuroscope along with the probe was forcibly withdrawn en bloc avulsing the frozen parietal pleura. Two to 3 samples were taken from each patient.
RESULTS
The diagnostic yield was 99% with cryobiopsy and 96% with forceps biopsy. The average specimen size through cryoprobe (13.2±6.7; range, 7 to 35 mm) was significantly larger than with the conventional forceps (6.8±3.3; range, 2 to 15 mm) (P<0.001), and no major complications were noted.
CONCLUSION
Cryobiopsy of the parietal pleura through the semirigid pleuroscope is a safe procedure with a very high diagnostic yield.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.