MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.clinbiomech.2005.08.009

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2005

Fatigue resistance analysis of tibial baseplate in total knee prosthesis--an in vitro biomechanical study.

Yu Tzai-Chiu, Huang Chang-Hung, Hsieh Chia-Hsun, Liau Jiann-Jong, Huang Chun-Hsiung, Cheng Cheng-Kung

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Tibial baseplates were occasionally reported with clinical fatigue failures. This study postulated that tibial baseplate of a specific mobile bearing design with a uniform thickness across the baseplate offers more fatigue resistance than the fixed-bearing design. Tibial baseplates of a fixed bearing and a mobile bearing design were fatigue-tested in vitro to study their fatigue resistance.

METHODS

Five samples of each design were tested under a sinusoidal loading between 90 N and 900 N at 30 Hz till failure or 10 million cycles. Experimental setup followed a standard published test method. Scanning electron microscope was used for inspecting the fracture surface of the failed baseplate.

FINDINGS: Two baseplates of fixed bearing design failed before 10 million cycles. Fatigue crack advancement marks were visible on the fractured surface of the failed samples. The fractured cross-section showed that the failure started near the end of the fin, it was likely due to the stress concentration as stress singularity existed at a point of sudden geometrical change. Five mobile bearing baseplates passed the test. Design of the tibial baseplate without fin structure and with a uniform thickness across the whole baseplate could help reducing the incidence of fatigue failure.

INTERPRETATION: The prosthesis survival rate was influenced by the long-term integrity of the metallic part of the prostheses such as the tibial baseplate. This study revealed that the tibial baseplate of a mobile bearing design with a uniform thickness provided better fatigue resistance than fixed bearing one. Standardized fatigue screening of the tibial baseplate was considered important in designing knee prostheses.

Для цитирования: Yu Tzai-Chiu, Huang Chang-Hung, Hsieh Chia-Hsun, Liau Jiann-Jong, Huang Chun-Hsiung, Cheng Cheng-Kung. Fatigue resistance analysis of tibial baseplate in total knee prosthesis--an in vitro biomechanical study.. Clinical biomechanics (Bristol, Avon). 2005. DOI: 10.1016/j.clinbiomech.2005.08.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.