DOI 10.3928/1081-597x-20031101-12
Relationship between cycloplegic and wavefront-derived refraction.
Аннотация
PURPOSE: To compare cycloplegic refractions to preoperative wavefront-derived refractions from three Hartmann-Shack-based wavefront devices.
METHODS
Ninety-nine eyes of 55 subjects were evaluated preoperatively using the LADARWave, WaveScan, and Zywave aberrometers. With cycloplegic refraction as the standard, the spherical equivalent refraction, sphere, and cylinder were compared to the wavefront-derived refractions using the three wavefront devices.
RESULTS
The LADARWave unit showed the highest correlation between wavefront-derived sphere and mean spherical equivalent refraction values relative to cycloplegic refraction, followed by Zywave and WaveScan. The WaveScan unit exhibited higher variability than the LADARWave or Zywave. Although Alcon Laboratories (LADARWave) and Bausch & Lomb Surgical (Zywave) recommend dilation, VISX specifically recommends against dilation in their WaveScan unit. All devices showed high cylinder error vectors when the cycloplegic cylinder was greater than 1.00 D.
CONCLUSION
Although wavefront-derived refractions appeared to correlate well to cycloplegic refractions, further research is required to reduce variability and improve the calculation of cylinder terms.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.