Insertion torque pullout strength relationship of soft tissue tendon graft tibia tunnel fixation with a bioabsorbable interference screw.

Nyland John, Kocabey Yavuz, Caborn David N M

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The purpose of this study was to evaluate the relationship between insertion torque and the biomechanical characteristics of soft tissue tendon graft tibia fixation with a bioabsorbable interference screw.

TYPE OF STUDY: Biomechanical study.

METHODS

Doubled tibialis anterior allografts (n = 20, length = 100 mm, diameter = 9 mm) prepared for anterior cruciate ligament (ACL) reconstructions were divided into 3 groups. Maximum insertion torque was measured as grafts were fixed by the same surgeon with a 10-mm diameter, 35-mm long screw in a 7-mm diameter tunnel (dilated to 9 mm) created in either 0.16 or 0.32 g/cm3 dense synthetic bone or cadaveric tibiae (0.84 +/- 0.15 g/cm2). Tensile testing to construct failure was then performed.

RESULTS

All constructs failed by tunnel pullout. Higher density synthetic bone had greater load at failure than lower density synthetic bone (486.6 +/- 83.8 v 271.6 +/- 52.6 N, P <.0001) and a slightly greater proportion of variance could be explained in synthetic bone (R2 = 0.74, P <.0001) than in cadaveric tibiae (R2 = 0.64, P =.017). Higher density synthetic bone displayed greater stiffness than lower density synthetic bone (68.1 +/- 21.6 v 37.7 +/- 11.2 N/mm, P =.012) and a higher proportion of variance could be explained in cadaveric tibiae (R2 = 0.72, P =.008) than in synthetic bone (R2 = 0.48, P =.012). Greater insertion torque levels were seen in higher density synthetic bone (2.97 +/- 0.35 v 1.24 +/- 0.06 Nm, P <.0001) than in lower density synthetic bone, and a higher proportion of variance could be explained in synthetic bone (R2 = 0.94, P <.0001) than in cadaveric tibiae (R2 = 0.74, P =.006).

CONCLUSIONS

Strong relationships were evident between insertion torque, bone mineral density (BMD) and load to failure across constructs. Differences between the synthetic bone groups and cadaveric tibiae, however, suggest the influence of mixed cortical and cancellous bone, nonuniform structural patterning, and differing BMD levels and distributions in the cadaveric tibiae.

CLINICAL RELEVANCE: Soft tissue graft tibial tunnel fixation is directly related to insertion torque and BMD. Synthetic models with mixed bone characteristics are needed to more accurately depict human tibiae during fixation testing.

Для цитирования: Nyland John, Kocabey Yavuz, Caborn David N M. Insertion torque pullout strength relationship of soft tissue tendon graft tibia tunnel fixation with a bioabsorbable interference screw.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2004. DOI: 10.1016/j.arthro.2004.01.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.