MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-2008-1027671

Медицинская библиотека · RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2008

[Ultrafast MR sialography: comparison of two coil systems including an alternative surface coil].

Weber T F, Cramer M C, Aldefeld D, Weiss F, Petersen K U, Reitmeier F, Jaehne M, Adam G, Habermann C R

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the visualization of salivary ducts by ultrafast magnetic resonance sialography (MR sialography) using an alternative surface coil and a conventional head-neck coil.

MATERIALS AND

METHODS

In 11 healthy volunteers, ultrafast MR sialography (single shot turbo spin echo; acquisition time, 2.8 s) was performed before and after oral application of a sialogogue. Each subject received examinations with both a bilateral surface coil (SC) and a conventional head-neck coil (HNC). The depiction of parotideal and submandibular duct systems was graded from 1 to 5 (5 = poor) by four independent radiologists. ANOVA served for statistical analysis of duct rankings, and interobserver variability was determined by Intraclass Correlation Coefficient (ICC).

RESULTS

With an excellent ICC of 0.96, both coil systems offered symmetric visualization of salivary ducts, and the image quality increased after oral application of sialogogue (p < 0.001). In total, the overall rating was worse for SC than for HNC (2.13 +/- 1.24 vs. 1.45 +/- 0.65, p < 0.001). SC was especially inferior in depiction of submandibular and extraglandular duct components compared to HNC (p < 0.001).

CONCLUSION

Most notably due to the reduced visualization of extraglandular and submandibular ducts, the specific surface coil used in this study was inferior in image quality and does not constitute a reasonable alternative to conventional coil systems.

Для цитирования: Weber T F, Cramer M C, Aldefeld D, Weiss F, Petersen K U, Reitmeier F, Jaehne M, Adam G, Habermann C R. [Ultrafast MR sialography: comparison of two coil systems including an alternative surface coil].. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2008. DOI: 10.1055/s-2008-1027671.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.