Evaluation of different low-dose multidetector CT and cone beam CT protocols in maxillary sinus imaging: part I-an in vitro study.

Almashraqi Abeer A, Ahmed Eman A, Mohamed Nashwa S, Barngkgei Imad H, Elsherbini Nader A, Halboub Esam S

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

This in vitro study aimed to assess radiation dose and image quality of different low-dose multidetector CT (MDCT) and CBCT imaging protocols in comparison with the standard MDCT protocol for maxillary sinus imaging.

METHODS

Effective dose (E) and image quality of 10 MDCT (changing effective milliampere second starting from 141.3 EmAs to 20 EmAs) and 3 CBCT protocols (changing milliampere second and voxel size) were assessed throughout scanning an anthropomorphic head and neck Alderson Rando phantom. E values were calculated using thermoluminescent dosemeters (TLDs) fixed at 6 sensitive organs (14 sites) on the Rando phantom. Image quality was assessed objectively (by calculating the standard deviation values of the radiographic density of water) and subjectively (by assessing the diagnostic image quality using a four-graded scale: 1 = very good, 2 = good, 3 = acceptable and 4 = unacceptable).

RESULTS

Two MDCT protocols (120 kV/32 EmA and 120 kV/25 EmA) had lower radiation doses with statistically significant differences (p < 0.001) compared with that of the standard MDCT protocol (120 kV/141.3 EmA), and they preserved a good diagnostic image quality. One CBCT protocol (120 kV/20 mA) had a reasonable radiation dose and good image quality. There were no statistically significant differences between the above-mentioned lower dose MDCT and CBCT protocols (p > 0.05) with respect to the radiation dose and image quality.

CONCLUSIONS

The low-dose MDCT and CBCT protocols are viable methods for maxillary sinus examination as evaluated using the above-mentioned phantom that yield a good diagnostic image quality using E approximately 7 and 11 times lower than that of the standard MDCT, respectively. These findings were evaluated in the in vivo part of this project.

Для цитирования: Almashraqi Abeer A, Ahmed Eman A, Mohamed Nashwa S, Barngkgei Imad H, Elsherbini Nader A, Halboub Esam S. Evaluation of different low-dose multidetector CT and cone beam CT protocols in maxillary sinus imaging: part I-an in vitro study.. Dento maxillo facial radiology. 2017. DOI: 10.1259/dmfr.20160323.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.