Novel Colorimetric Capnography for Confirmation of Thoracostomy Placement in a Porcine Model of Pneumothorax.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Topper Gena V, Buonasorte Marlena, Thange Lamis, Miles Avery, Armento Isabella, Maitha John C, Hunter Krystal, Egodage Tanya
Аннотация
INTRODUCTION: Tension pneumothorax is a rapidly fatal but pre ventable cause of death. Needle thoracostomy remains the first-line intervention, yet misplaced catheters fail to relieve pressure and may prolong hypoxia, resulting in cardiovascular compromise and potentially causing injury. This study evaluated the utility of a novel colorimetric capnography device (Capnospot™) designed to provide visual confirmation of pleural entry during decompression.
METHODS: Two swine models of tension pneumothorax were created using transdiaphragmatic CO2 insufflation. After each induction, either needle or pigtail thoracostomy was performed. Time to Capnospot® color change was compared with ultrasound and radiographic confirmation using the Wilcoxon signed-rank test.
RESULTS: Twenty-four thoracostomies were performed. Time to colorimetric capnography change for needle thoracostomy was shorter than time to ultrasound confirmation (1,030 vs. 7,030 milliseconds (ms), P=.004). Time to color change was shorter than both the time to X-ray confirmation (1,030 vs. 9,435ms, P=.002) and ultrasound confirmation for pigtail thoracostomy placement (355 vs. 22,355ms, P=.002).
CONCLUSIONS: Colorimetric capnography provided rapid, reliable confirmation of thoracostomy placement, outperforming both ultrasound and X-ray in speed and accuracy. This technology may streamline decompression in both prehospital and in-hospital settings, reducing complications and delays in life-saving care.
METHODS: Two swine models of tension pneumothorax were created using transdiaphragmatic CO2 insufflation. After each induction, either needle or pigtail thoracostomy was performed. Time to Capnospot® color change was compared with ultrasound and radiographic confirmation using the Wilcoxon signed-rank test.
RESULTS: Twenty-four thoracostomies were performed. Time to colorimetric capnography change for needle thoracostomy was shorter than time to ultrasound confirmation (1,030 vs. 7,030 milliseconds (ms), P=.004). Time to color change was shorter than both the time to X-ray confirmation (1,030 vs. 9,435ms, P=.002) and ultrasound confirmation for pigtail thoracostomy placement (355 vs. 22,355ms, P=.002).
CONCLUSIONS: Colorimetric capnography provided rapid, reliable confirmation of thoracostomy placement, outperforming both ultrasound and X-ray in speed and accuracy. This technology may streamline decompression in both prehospital and in-hospital settings, reducing complications and delays in life-saving care.