Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp.

Gao Xinyu, Zhang Chao, Wang Yun, Lu Xiaomei, Yu Shenghui, Zheng Wenjie, Lin Zehong, Wang Yining, Xue Huadan, Zhang Daming, Jin Zhengyu

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study quantitatively evaluated the effects of tube voltage, CTDIvol, energy level, and iodine concentration on the performance of lower-extremity vascular imaging using a second-generation dual-layer spectral CT (DSCT). We further assessed the potential for reducing radiation dose and contrast agent use without compromising image quality.

METHODS

A spectral CT phantom with tissue-equivalent inserts and varying iodine concentrations (4, 8, 12, and 16 mgI/mL) was scanned at different tube voltages (100/120 kVp), CTDIvol levels (2.5, 5, 10 mGy), and virtual monoenergetic levels (40, 70, 100 keV), alongside conventional imaging. Vessel diameters of 2.5, 4, and 5 mm were evaluated. Objective image quality was assessed via CT attenuation, SD, SNR, and CNR. Two radiologists used a 5-point Likert scale to assess the subjective image quality.

RESULTS

No significant differences were observed in subjective or objective scores between tube voltages or across CTDIvol levels (p > 0.05). Low-energy 40 keV images yielded comparable quality to 70 keV and conventional images. At 40 keV, ultra-low iodine concentrations (4 mgI/mL) achieved image quality similar to higher concentrations (12/16 mgI/mL) in conventional scans. The combination of 100 kVp, 2.5 mGy, and 40 keV provided superior image quality relative to conventional CT.

CONCLUSIONS

Second-generation DSCT at 100 kVp provides high-accuracy lower-limb vascular imaging, even at low radiation doses and contrast concentrations, supporting its potential for clinical dose-reduction strategies.

Для цитирования: Gao Xinyu, Zhang Chao, Wang Yun, Lu Xiaomei, Yu Shenghui, Zheng Wenjie, Lin Zehong, Wang Yining, Xue Huadan, Zhang Daming, Jin Zhengyu. Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp.. BMC medical imaging. 2025. DOI: 10.1186/s12880-025-02047-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.