Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans.

Buchhave E Ø, Rasmussen C D, Lentz R, Frederiksen E H, Kipp J O, Leftes M Ø

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose (E), when positioning the arms below shoulder level in CT-TAP.

METHODS

An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate E. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.

RESULTS

Significant differences in E between all arm positions were seen (p = 0.005). The lowest E was obtained in position C, reducing E by 8.42%. Position A provided the best image quality but the highest E.

CONCLUSION

This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest E. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.

IMPLICATION FOR PRACTICE: This study illustrates tendencies for objective image quality and E when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.

Для цитирования: Buchhave E Ø, Rasmussen C D, Lentz R, Frederiksen E H, Kipp J O, Leftes M Ø. Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans.. Radiography (London, England : 1995). 2024. DOI: 10.1016/j.radi.2024.04.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.