Clinical gastroenterology and hepatology : the official clinical practice journal of the American Gastroenterological Association · 2024
Accuracy of Computer-aided Diagnosis in Colonoscopy Varies According to Polyp Location: A Systematic Review and Meta-analysis.
СтатьяСистематический обзорМетаанализНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rizkala Tommy, Hassan Cesare, Mori Yuichi, Spadaccini Marco, Antonelli Giulio, Dekker Evelien, Houwen Britt B S L, Pech Oliver, Baumer Sebastian, Rondonotti Emanuele, Radaelli Franco, Li James Weiquan, von Renteln Daniel, Misawa Masashi, Facciorusso Antonio, Maselli Roberta, Carrara Silvia, Fugazza Alessandro, Capogreco Antonio, Khalaf Kareem, Patel Harsh, Sharma Prateek, Rex Douglas, Repici Alessandro
Аннотация
BACKGROUND & AIMS: Computer-aided diagnosis (CADx) assists endoscopists in differentiating between neoplastic and non-neoplastic polyps during colonoscopy. This study aimed to evaluate the impact of polyp location (proximal vs. distal colon) on the diagnostic performance of CADx for ≤5 mm polyps.
METHODS: We searched for studies evaluating the performance of real-time CADx alone (ie, independently of endoscopist judgement) for predicting the histology of colorectal polyps ≤5 mm. The primary endpoints were CADx sensitivity and specificity in the proximal and distal colon. Secondary outcomes were the negative predictive value (NPV), positive predictive value (PPV), and the accuracy of the CADx alone. Distal colon was limited to the rectum and sigmoid.
RESULTS: We included 11 studies for analysis with a total of 7782 polyps ≤5 mm. CADx specificity was significantly lower in the proximal colon compared with the distal colon (62% vs 85%; risk ratio (RR), 0.74; 95% confidence interval [CI], 0.72-0.84). Conversely, sensitivity was similar (89% vs 87%); RR, 1.00; 95% CI, 0.97-1.03). The NPV (64% vs 93%; RR, 0.71; 95% CI, 0.64-0.79) and accuracy (81% vs 86%; RR, 0.95; 95% CI, 0.91-0.99) were significantly lower in the proximal than distal colon, whereas PPV was higher in the proximal colon (87% vs 76%; RR, 1.11; 95% CI, 1.06-1.17).
CONCLUSION: The diagnostic performance of CADx for polyps in the proximal colon is inadequate, exhibiting significantly lower specificity compared with its performance for distal polyps. Although current CADx systems are suitable for use in the distal colon, they should not be employed for proximal polyps until more performant systems are developed specifically for these lesions.
METHODS: We searched for studies evaluating the performance of real-time CADx alone (ie, independently of endoscopist judgement) for predicting the histology of colorectal polyps ≤5 mm. The primary endpoints were CADx sensitivity and specificity in the proximal and distal colon. Secondary outcomes were the negative predictive value (NPV), positive predictive value (PPV), and the accuracy of the CADx alone. Distal colon was limited to the rectum and sigmoid.
RESULTS: We included 11 studies for analysis with a total of 7782 polyps ≤5 mm. CADx specificity was significantly lower in the proximal colon compared with the distal colon (62% vs 85%; risk ratio (RR), 0.74; 95% confidence interval [CI], 0.72-0.84). Conversely, sensitivity was similar (89% vs 87%); RR, 1.00; 95% CI, 0.97-1.03). The NPV (64% vs 93%; RR, 0.71; 95% CI, 0.64-0.79) and accuracy (81% vs 86%; RR, 0.95; 95% CI, 0.91-0.99) were significantly lower in the proximal than distal colon, whereas PPV was higher in the proximal colon (87% vs 76%; RR, 1.11; 95% CI, 1.06-1.17).
CONCLUSION: The diagnostic performance of CADx for polyps in the proximal colon is inadequate, exhibiting significantly lower specificity compared with its performance for distal polyps. Although current CADx systems are suitable for use in the distal colon, they should not be employed for proximal polyps until more performant systems are developed specifically for these lesions.