Risk of airway fire with the use of KTP laser and high flow humidified oxygen delivery in a laryngeal surgery model.

Huang Lucy, Badenoch Adam, Vermeulen Marthinus, Ullah Shahid, Woods Charmaine, Athanasiadis Theodore, Ooi Eng Hooi

СтатьяПеревод пока не подготовлен

Аннотация

Airway surgery presents a unique environment for operating room fire to occur. This study aims to explore the factors of combustion when using KTP laser with high flow oxygen in an ex-vivo model. The variables tested were varying tissue type, tissue condition, oxygen concentration, laser setting, and smoke evacuation in a stainless-steel model. Outcome measures were time of lasing to the first spark and/or flame. A multivariate Cox proportional hazard model was used to determine the risk of spark and flame across the different risk factors. For every 10% increase in oxygen concentration above 60% the risk of flame increased by a factor of 2.3. Continuous laser setting at 2.6 W increased the risk by a factor of 72.8. The risk of lasing adipose tissue is 7.3 times higher than that of muscle. Charred tissue increases the risk of flame by a factor of 92.8. Flame occurred without a preceding spark 93.6% of the time. Using KTP laser in the pulsed mode with low wattages, minimising lasing time, reducing the oxygen concentration and avoiding lasing adipose or charred tissue produce a relatively low estimated risk of spark or flame.

Для цитирования: Huang Lucy, Badenoch Adam, Vermeulen Marthinus, Ullah Shahid, Woods Charmaine, Athanasiadis Theodore, Ooi Eng Hooi. Risk of airway fire with the use of KTP laser and high flow humidified oxygen delivery in a laryngeal surgery model.. Scientific reports. 2022. DOI: 10.1038/s41598-021-04636-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.