Proprieties of adhesive surface arrays to thyroid cartilage for recurrent laryngeal nerve monitoring.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Zhao Yishen, Zhang Daqi, Zhou Le, Li Shijie, Wang Tie, Li Fang, Han Yujia, Dionigi Gianlorenzo, Sun Hui

Аннотация

BACKGROUND: To investigate the optimal placement and size of adhesive thyroid cartilage electrodes (ATCEs) placed on the thyroid cartilage in porcine models for evaluating recurrent laryngeal nerve function compared with the other 3 recording electrode types.

METHODS: Four Meishan piglets were used for this study. The electromyogram (EMG) signal stability was detected by intraoperative nerve monitoring (IONM). Best location and size of ATCEs were detected, and the EMG signals from each were compared. Latency data were tested by stimulating the different nerve points, and 6 manipulations of the trachea and endotracheal tube (ET) were applied to test the stability of EMG signals. ET electrodes, needle electrodes, and transcutaneous recording electrodes were simultaneously tested with comparison to ATCEs.

RESULTS: The optimal placement locations and sizes of ATCEs were determined. The amplitudes and latencies recorded from the ATCEs were consistent with those of the ET electrodes. More anti-interference was observed with the ATCEs than with the ET electrodes in surgical manipulations. ATCEs could be used during intermittent and continuous monitoring in similar fashion to ET electrodes and needle electrodes.

CONCLUSIONS: ATCEs had consistent monitoring function with ET electrodes and needle electrodes. Feasibility, EMG stability, and optimal location and size of ATCEs for IONM were resolved. The EMG profiles from the ATCEs were more stable during surgical manipulations.