Influence of the urethral pressure transducer in measuring Valsalva leak point pressure in women undergoing multichannel urodynamic testing.
Аннотация
AIMS: To determine if the air-charged urethral sensor balloon currently used in urodynamic testing (UDS) significantly impacts Valsalva leak point pressure (VLPP) measurements.
METHODS
This is a prospective cohort study of women undergoing UDS at an academic institution. VLPPs were obtained at 150 mL and urodynamic capacity with and without the urethral pressure sensor in the urethra. VLPP measurements were analyzed using a Wilcoxon signed-rank test. Median and interquartile range are presented.
RESULTS
Sixty-three patients were enrolled in the study, 53 were included in the primary analysis. The mean age of the subjects was 56.2 ± 12.1 years. Nine patients (16%) solely leaked when the balloon was not present in the urethra either with cough or during VLPP measurement. At both 150 mL and urodynamic capacity, when VLPP testing was performed, there was a significant difference (cmH2 O) between the control and intervention values, (76.2 [55.0, 97.0] vs 68.8 [46.3, 93.3], P = .0012; 79.3 [53, 96.5] vs 72.5 [50.8, 92.3], P = .04). There was also a statistically significant difference between the control and intervention values for the lowest leak value at 150 mL and capacity (70.5 [51, 94.5] vs 60.0 [40, 88] P = .002; 73.5 [49.5, 91.5] vs 61 [45, 88], P = .017).
CONCLUSIONS
The higher VLPPs obtained with the urethral balloon in place indicate that the balloon may be the cause of falsely elevated VLPPs during urodynamic testing. Additionally, the balloon may mask a diagnosis of stress urinary incontinence in some patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.