Comparison of the endoscopic surgical monitoring system and blood parameters for evaluating irrigation fluid absorption during bipolar plasmakinetic transurethral resection of the prostate: a prospective observational study.

Jia Haitao, Zheng Qihui, Bai Jiangtao, Zhang Luyang, Jin Qi, Wang Guilin, Guo Ting, Mi Jun, Zhang Yunxin, Tian Junqiang, Yang Li, Wang Zhiping, Dong Zhilong

СтатьяНаблюдательное исследованиеСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Irrigation fluid absorption is an unpredictable complication during transurethral resection of the prostate (TURP). This prospective study aimed to compare the Endoscopic Surgical Monitoring System (ESMS) and blood parameters for evaluating fluid absorption volume in patients undergoing bipolar plasmakinetic TURP.

METHODS

Bipolar TURP was performed using 0.9% saline for bladder irrigation. The ESMS measured fluid absorption volumes in real time. In addition, arterial blood samples were collected for blood gas analysis immediately before irrigation and after surgery, with the absorption volume calculated using equations based on changes in plasma sodium, chloride, potassium, calcium, and hemoglobin concentrations.

RESULTS

A total of 420 patients (mean age 69.8 ± 7.7 years) were included. All calculated volumes based on blood parameters were higher than the ESMS-measured volume. The volume calculated from plasma chloride concentration changes showed the closest agreement with the ESMS volume (867.5 ± 748.9 vs. 843.3 ± 722.1 mL, P = 0.416), with a mean bias of 14.4 mL and 95% agreement limits of -385.7 to 414.5 mL. Regression analysis demonstrated a strong linear relationship between plasma chloride elevation and actual absorption volumes (R²=0.776, P = 0.001). However, further analysis revealed a mean percentage bias of + 19%, indicating a systematic overestimation tendency in blood chloride method. Furthermore, a clinically significant discrepancy (> 300 mL) between the blood chloride method and ESMS was associated with altered clinical management, including increased diuretic use (36.7% vs. 15.0%, P < 0.001) and limitation of intravenous fluids (33.3% vs. 17.5%, P = 0.004).

CONCLUSIONS

Among the blood parameters, plasma chloride provides the most accurate estimation of irrigation fluid absorption and can serve as a practical hematologic indicator when ESMS is unavailable. However, clinicians should be aware of its inherent overestimation tendency and the potential for clinically significant discrepancies to impact patient management. TRIAL REGISTRATION: This study was prospectively registered in the Chinese Clinical Trial Registry on December 25, 2023 ( www.chictr.org.cn ; Registration Number: ChiCTR2300079095).

Для цитирования: Jia Haitao, Zheng Qihui, Bai Jiangtao, Zhang Luyang, Jin Qi, Wang Guilin, Guo Ting, Mi Jun, Zhang Yunxin, Tian Junqiang, Yang Li, Wang Zhiping, Dong Zhilong. Comparison of the endoscopic surgical monitoring system and blood parameters for evaluating irrigation fluid absorption during bipolar plasmakinetic transurethral resection of the prostate: a prospective observational study.. BMC anesthesiology. 2026. DOI: 10.1186/s12871-026-03652-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.