Evaluation of thyroid radiation dose during abdominal Computed Tomography procedures and dose reduction effectiveness of thyroid shielding.
Аннотация
INTRODUCTION
During abdominal Computed Tomography (CT) studies, vicinity organs receive a dose from scatter radiation. The thyroid is considered an organ at greater risk due to high radiosensitivity.
METHODS
The primary objective of this study was to determine the entrances surface dose (ESD) to the thyroid during abdominal CT studies and to evaluate the efficiency of dose reduction by lead shielding. The calibrated thermoluminescence dosimeter (TLD) chips were used to measure the ESD during 180 contrast-enhanced (CE) and non-contrast-enhanced (NC) abdominal CT studies in the presence and absence of lead shielding.
RESULTS
Thyroid shielding reduces the ESD by 72.3% (0.55 mGy), 86.5% (2.95 mGy) and 64.0% (2.24 mGy) during NC, 3-phase and 4-phase abdominal CT scans. Also, the patient height was identified as a parameter that inversely influenced the thyroid dose, proving that the taller patients receive less dose to the thyroid. Regardless, the scan parameters such as time and display field of view (DFOV) positively impact the thyroid dose.
CONCLUSION
Lead shielding can prevent the external scatter reaching the thyroid region by 64%-87% during the non-vicinity scans such as abdomen CT. However, the actual dose saving lies between 0.2% and 0.4%, compared to the total effective dose of the whole CT procedure.
IMPLICATIONS FOR PRACTICE: The thyroid shield can effectively reduce external scatter radiation during abdominal CT procedures. However, the dose saving is insignificant compared to the total effective dose from the whole examination. Therefore, the use of thyroid shielding should be carefully evaluated during CT abdomen procedures.