Digital image composition in long-leg radiography with a flat-panel detector: first clinical experiences.

Völk Marcus, Angele Peter, Hamer Okka, Feuerbach Stefan, Strotzer Michael

СтатьяПеревод пока не подготовлен

Аннотация

RATIONALE AND

OBJECTIVES

The purpose of this study was to evaluate the image quality of composed long-leg examinations with a large-area, flat-panel x-ray detector.

METHODS

Thirty-five consecutive patients were included in this study. All images were obtained with a kilovoltage setting identical with conventional radiographies of speed class 400; amperage values were reduced by 50% compared with standard dose. After acquisition, the images were transferred to a workstation where the whole image was reconstructed using a generalized correlation method. Images were presented to 3 observers. Examination quality was ranked on a 3-point scale: 1 = no manual adjustment necessary; 2 = composition required manual correction; 3 = no composition possible.

RESULTS

Patient rankings were 31/35 (88.6%) in category 1, 3/35 (8.6%) in category 2, and 1/35 (2.8%) in category 3 (primarily due to an application error).

CONCLUSIONS

The analysis of the first clinical examinations of long-leg radiographies with a 43 cm x 43 cm flat-panel detector demonstrates very good reliability of the digital image composition.

Для цитирования: Völk Marcus, Angele Peter, Hamer Okka, Feuerbach Stefan, Strotzer Michael. Digital image composition in long-leg radiography with a flat-panel detector: first clinical experiences.. Investigative radiology. 2003. DOI: 10.1097/01.rli.0000057028.63568.85.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.