Bioanalysis · 2025
Quantification of keratan sulfate in biofluids by validated HPCE-LIF: application to Morquio syndrome.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Galeotti Fabio, Zampini Lucia, Maccari Francesca, Monachesi Chiara, Padella Lucia, Santoro Lucia, Gabrielli Orazio, Volpi Nicola
Аннотация
BACKGROUND: Keratan sulfate (KS) is altered in several pathological conditions. We report the capillary electrophoresis-laser induced fluorescence separation and validation of KS-derived disaccharides Galβ(1-4)GlcNAc(6S) and Gal(6S)β(1-4)GlcNAc(6S) in human urine and plasma after keratanase II treatment and AMAC fluorotagging.
RESULTS: The calibration curves of the two disaccharides from 50 to 1000 ng showed an average correlation coefficient greater than 0.9970 and the calculated LOD and LOQ were 3 and 10 ng, respectively. The inter-day precision was ~13% and the inter-day accuracy was ~10% for both disaccharides. The % recovery ranged between 88% and 106% for 200, 500, and 800 ng of each disaccharide added to biofluids. Urine and plasma of Morquio A and B subjects and of patients submitted to ERT treatment were tested. Urinary KS was ~13-16-fold more abundant in MPSIVA and plasmatic KS was significantly ~2-3-fold more than controls. No differences were observed for the disaccharides ratio between Patients and not-affected Subjects.
CONCLUSION: The capillary electrophoresis separation of the two main KS disaccharides in human biofluids was found specific, sensitive, and accurate and suitable for application to Morquio syndrome early detection and progression as well as for other conditions in which KS is altered in content and structure.
RESULTS: The calibration curves of the two disaccharides from 50 to 1000 ng showed an average correlation coefficient greater than 0.9970 and the calculated LOD and LOQ were 3 and 10 ng, respectively. The inter-day precision was ~13% and the inter-day accuracy was ~10% for both disaccharides. The % recovery ranged between 88% and 106% for 200, 500, and 800 ng of each disaccharide added to biofluids. Urine and plasma of Morquio A and B subjects and of patients submitted to ERT treatment were tested. Urinary KS was ~13-16-fold more abundant in MPSIVA and plasmatic KS was significantly ~2-3-fold more than controls. No differences were observed for the disaccharides ratio between Patients and not-affected Subjects.
CONCLUSION: The capillary electrophoresis separation of the two main KS disaccharides in human biofluids was found specific, sensitive, and accurate and suitable for application to Morquio syndrome early detection and progression as well as for other conditions in which KS is altered in content and structure.