Laryngoscopic evaluation with the Airway Cam.

George Ronald B, McKeen Dolores, Law J Adam

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The Airway Cam is a head-mounted direct laryngoscopy video system which uses a prismatic sighting system that aligns with the operator's line of sight. This study evaluated intra- and inter-observer consistency in laryngoscopy grading comparing direct laryngoscopy to the laryngoscopy grade obtained with the Airway Cam.

METHODS

Twenty-seven patients receiving a general anesthetic for elective surgery had laryngoscopy performed by an anesthesiologist wearing the Airway Cam. Each video was duplicated, then randomized and reviewed in a blinded fashion by the original laryngoscopist and a second anesthesiologist. Intra- and inter-observer correlations were identified.

RESULTS

There was good intra-observer agreement of the Cormack-Lehane scale between direct laryngoscopy and laryngoscopy recorded with the Airway Cam (kappa = 0.63). The corresponding intra-observer correlation of the percentage of glottic opening score was strong at r = 0.83. There was good inter-observer agreement of the Cormack-Lehane scale between direct laryngoscopy and that observed by the second anesthesiologist during Airway Cam video review (kappa = 0.70). There was moderate correlation of the inter-observer percentage of glottic opening scores (r = 0.73).

CONCLUSION

This study validates that the view recorded by the Airway Cam reflects the view of the laryngoscopist. The Airway Cam may introduce an additional level of objectivity into airway management research and teaching.

Для цитирования: George Ronald B, McKeen Dolores, Law J Adam. Laryngoscopic evaluation with the Airway Cam.. Canadian journal of anaesthesia = Journal canadien d'anesthesie. 2006. DOI: 10.1007/bf03022626.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.