DOI 10.1016/j.urology.2005.12.022
Evaluation of a novel knotless self-anchoring suture material for urinary tract reconstruction.
Аннотация
OBJECTIVES
To evaluate a novel prototype self-anchoring suture (SAS) material that incorporates unidirectional "barbs" designed to grip and approximate tissue without the need for suture ligation for urinary tract reconstructive procedures.
METHODS
The in vitro failure strength of various porcine tissue approximations with SAS was compared with the strength of similar approximations with standard ligated suture. For in vivo analysis, 3 pigs underwent laparoscopic pyeloplasty with SAS on one side and 2-0 Vicryl suture on the contralateral side. Nine pigs underwent bladder neck anastomoses (BNA) performed with SAS (n = 3) and standard ligated interrupted suturing (n = 3) or standard ligated running suturing (n = 3). The operative times were compared. One week after surgery, the integrity of each anastomosis was evaluated radiographically and histopathologically.
RESULTS
The in vitro analysis revealed no statistically different (P = 0.22) failure forces between the approximations performed with SAS (13.2 +/- 2.6 N) and standard suture (14.1 +/- 3.1 N). The in vitro analysis revealed no difference in operative times for the pyeloplasties (P = 0.72) or BNA (P = 0.31). None of the 1-week postoperative radiographic studies revealed extravasation. The histopathologic analysis revealed more fibrotic reaction associated with the SAS BNA procedures (P <0.01).
CONCLUSIONS
Self-anchoring suture secures tissue approximations at loads equivalent to tissue approximations with standard tied suture. Self-anchoring suture obviates the need for knot tying and provides a watertight anastomosis. With laparoscopic knot tying experience, anastomotic time with SAS and standard suture do not differ. Self-anchoring suture might induce more fibrosis. Long-term follow-up evaluation will be required before clinical application.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.