DOI 10.1016/s0302-2838(02)00281-6
Externally coated ureteral metallic stents: an unfavorable clinical experience.
Аннотация
OBJECTIVE
The purpose of the present study was to evaluate the use of externally coated stents in patients with malignant ureteral obstruction.
MATERIALS AND
METHODS
We have prospectively evaluated 16 patients, 10 men and 6 women, with malignant ureteral obstruction treated successfully by placement of Passager metal stents (Boston Scientific, Natick, MA, USA) bypassing the stricture. Mean patient age was 65.6 years (range 62-78 years). Ureteral patency was confirmed 24 and 48 hours by injection of contrast material through the nephrostomy tube, and after patency confirmation the nephrostomy catheter was removed.
RESULTS
All stents were positioned successfully, and the postoperative course was uneventful. In 13 cases (81.2%) the prostheses finally migrated into the bladder hindering overall ureteral patency (mean time of migration: 1.5 months). Patency was achieved in the remaining ureters (n=3), during the follow-up period (mean: 8 months, range 6-16 months), without any need for further intervention.
CONCLUSION
The inappropriate anchorage and the increased ureteral peristalsis are the main causes of migration towards the bladder, thus, minimizing the usefulness of this stent for the treatment of ureteral strictures.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.