Extracorporeal photopheresis with permanent subcutaneous right atrial catheters.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Ständer Hartmut, Neugebauer Frank, Schneider Stefan W, Luger Thomas A, Schiller Meinhard
Аннотация
BACKGROUND: Adequate peripheral venous access is crucial for successful extracorporeal photopheresis (ECP). As this approach is not always feasible in older patients and patients with graft-versus-host disease, central venous catheters play an increasing role in providing long-term vascular access for ECP.However, not all catheters are able to deliver the minimum flow rate of 7 ml/min for ECP.
PATIENTS AND METHODS: Eight different permanent subcutaneous right atrial catheters were connected in vitro to the UVAR-XTS photopheresis system and median flow rates were determined. In addition, in vivo flow rates of patients who received ECP, using either peripheral or central venous access, were determined.
RESULTS: Hemodialysis catheters with an internal diameter of 2.0 or 1.5 x 3.5 mm and a length up to 48 cm provided in vitro flow rates of 27-28 ml/min, almost identical to a peripheral access needle. Central venous catheters with a length of over 90 cm reached flow rates below 7 ml/min and are impractical for ECP. The analysis of 308 ECP collection cycles with peripheral vascular access revealed an average flow rate of 31.5 +/- 6.4 ml/min. Only permanent subcutaneous right atrial catheters made for hemodialysis provided similar flow rates (Quinton PermCath Dual Lumen) (33.7 +/- 4.7 ml/min, n = 198).
CONCLUSIONS: Permanent subcutaneous hemodialysis catheters with a length of maximally 48 cm achieve optimal flow rates for ECP. They represent therefore the central venous access of choice in patients with inadequate peripheral vascular access.
PATIENTS AND METHODS: Eight different permanent subcutaneous right atrial catheters were connected in vitro to the UVAR-XTS photopheresis system and median flow rates were determined. In addition, in vivo flow rates of patients who received ECP, using either peripheral or central venous access, were determined.
RESULTS: Hemodialysis catheters with an internal diameter of 2.0 or 1.5 x 3.5 mm and a length up to 48 cm provided in vitro flow rates of 27-28 ml/min, almost identical to a peripheral access needle. Central venous catheters with a length of over 90 cm reached flow rates below 7 ml/min and are impractical for ECP. The analysis of 308 ECP collection cycles with peripheral vascular access revealed an average flow rate of 31.5 +/- 6.4 ml/min. Only permanent subcutaneous right atrial catheters made for hemodialysis provided similar flow rates (Quinton PermCath Dual Lumen) (33.7 +/- 4.7 ml/min, n = 198).
CONCLUSIONS: Permanent subcutaneous hemodialysis catheters with a length of maximally 48 cm achieve optimal flow rates for ECP. They represent therefore the central venous access of choice in patients with inadequate peripheral vascular access.