MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/00016489.2021.1907444

Медицинская библиотека · Acta oto-laryngologica · 2021

Endoscopic visualization for tracheoesophageal puncture tract sizing.

Evangelista Lisa M, Andrews Tess, Bayoumi Ahmed, Kuhn Maggie A, Nativ-Zeltzer Nogah, Belafsky Peter C

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurate tracheoesophageal prosthesis (TEP) size is essential for optimal functioning. Current sizing techniques do not afford direct visualization. The clinical utility of direct visualization with unsedated transnasal esophagoscopy (TNE) to improve TEP sizing accuracy has not been established.

AIMS/

OBJECTIVES

The purpose of this investigation was to determine if endoscopic visualization with TNE during TEP fitting improves sizing accuracy.

MATERIAL AND

METHODS

Participants undergoing secondary TEP placement had their TEP sized without visualization and with visualization during TNE. The effect of endoscopic visualization on TEP sizing was assessed.

RESULTS

The mean age (SD) of the cohort (N = 15) was 61.20 (±10.19) years. The percent disagreement between non-visualized and endoscopic-visualized puncture tract measurement was 60%. The mean difference in puncture tract length with visualization compared to without visualization was 1.75 mm (±1.91). Discordance between non-visualized and endoscopic-visualized puncture tract length was worse for persons with a history of radiation (66.7%) than without radiation (33.3%) (p < .05). The overall test agreement between endoscopic and non-endoscopic sizing techniques was moderate (Cohen's kappa coefficient = 0.254; (p < .05).

CONCLUSION AND SIGNIFICANCE: The percent disagreement between non-visualized and endoscopic-visualized TEP sizing is high (60%). The data suggest that endoscopic visualization with TNE improves sizing accuracy during TEP placement.

Для цитирования: Evangelista Lisa M, Andrews Tess, Bayoumi Ahmed, Kuhn Maggie A, Nativ-Zeltzer Nogah, Belafsky Peter C. Endoscopic visualization for tracheoesophageal puncture tract sizing.. Acta oto-laryngologica. 2021. DOI: 10.1080/00016489.2021.1907444.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.