Rapid visual identification of PCR amplified nucleic acids by centrifugal gel separation: Potential use for molecular point-of-care tests.

Hwang Sang-Hyun, Kim Dong-Eun, Im Ji-Hyun, Kang Su-Jin, Lee Do-Hoon, Son Sang Jun

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Recently, nucleic acid amplification and detection techniques have progressed based on advances in in microfluidics, microelectronics, and optical systems. Nucleic acids amplification based point-of-care test (POCT) in resource-limited settings requires simple visual detection methods. Several biosensing methods including lateral flow immunoassays (LFIA) were previously used to visually detect nucleic acids. However, prolonged assay time, several washing steps, and a need for specific antibodies limited their use. Here we developed a novel, rapid method to visualize amplified nucleic acids with naked eyes in clinical samples. First, we optimized conditions based on separation using very low centrifugal force and a density medium to detect human papillomavirus (HPV)-16 DNA in cervical specimens. After DNA extraction, HPV16 PCR was performed with biotin-labeled forward primer and Cy3-labeled reverse primer. PCR amplicon was mixed with streptavidin-magnetic beads, introduced into the density medium. After two-minute centrifugation, the result was visually identified. This system showed identical results with commercial HPV real-time PCR for 30 clinical samples and could detect up to 10(2)copies/mL of HPV DNA without any optical instruments. This robust and sensitive visual detection system is suitable for non-specialist personnel and point-of-care diagnosis in low-resource settings.

Для цитирования: Hwang Sang-Hyun, Kim Dong-Eun, Im Ji-Hyun, Kang Su-Jin, Lee Do-Hoon, Son Sang Jun. Rapid visual identification of PCR amplified nucleic acids by centrifugal gel separation: Potential use for molecular point-of-care tests.. Biosensors & bioelectronics. 2016. DOI: 10.1016/j.bios.2016.01.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.