MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1167/tvst.13.5.19

Медицинская библиотека · Translational vision science & technology · 2024

Design and Biocompatibility of a Novel, Flexible Artificial Cornea.

Li Gavin, Aldave Anthony J, Amescua Guillermo, Colby Kathryn A, Cortina Maria S, de la Cruz Jose, Parel Jean-Marie A, Schmiedel Thomas B, Akpek Esen Karamursel

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: We sought to introduce the materials, design, and biocompatibility of a flexible and suturable artificial corneal device.

METHODS

Single-piece, fully synthetic, optic-skirt design devices were made from compact perfluoroalkoxy alkane. The skirt and the optic wall surfaces were lined with a porous tissue ingrowth material using expanded polytetrafluoroethylene. Full-thickness macroapertures around the skirt perimeter were placed to facilitate nutrition of the recipient cornea. Material properties including the skirt's modulus of elasticity and bending stiffness, optic light transmission, wetting behavior, topical drug penetrance, and degradation profile were evaluated.

RESULTS

The final prototype suitable for human use has a transparent optic with a diameter of 4.60 mm anteriorly, 4.28 mm posteriorly, and a skirt outer diameter of 6.8 mm. The biomechanical and optical properties of the device closely align with the native human cornea with an average normalized device skirt-bending stiffness of 4.7 kPa·mm4 and light transmission in the visible spectrum ranging between 92% and 96%. No optical damage was seen in the 36 devices tested in fouling experiments. No significant difference was observed in topical drug penetrance into the anterior chamber of the device implanted eye compared with the naïve rabbit eye.

CONCLUSIONS

The flexibility and biocompatibility of our artificial cornea device may offer enhanced tissue integration and decreased inflammation, leading to improved retention compared with rigid keratoprosthesis designs.

TRANSLATIONAL RELEVANCE: We have developed a fully synthetic, flexible, suturable, optic-skirt design prototype artificial cornea that is ready to be tested in early human feasibility studies.

Для цитирования: Li Gavin, Aldave Anthony J, Amescua Guillermo, Colby Kathryn A, Cortina Maria S, de la Cruz Jose, Parel Jean-Marie A, Schmiedel Thomas B, Akpek Esen Karamursel. Design and Biocompatibility of a Novel, Flexible Artificial Cornea.. Translational vision science & technology. 2024. DOI: 10.1167/tvst.13.5.19.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.