DOI 10.17116/oftalma202414002243

Медицинская библиотека · Vestnik oftalmologii · 2024

[Influence of corneal curvature and peripheral thickness on tonometry readings].

Antonov A A, Volzhanin A V, Klinicheva E A, Karpilova M A

Аннотация на английскомСтатьяОригинал на русском

Аннотация

PURPOSE: This study investigates the influence of peripheral corneal thickness (PCT) and its curvature on tonometry readings.

MATERIAL AND

METHODS

The study included 49 patients (49 eyes) who were indicated for glaucoma surgery. Using bidirectional applanation tonometry, the following parameters were obtained: IOPcc, IOPg - intraocular pressure (IOP) corrected for corneal compensation, taken as the most reliable indicator; IOP converted to Goldmann measurement, taken as the result of applanation tonometry, ΔIOP (IOPcc-IOPg), CH and CRF (corneal hysteresis and corneal resistance factor). During corneal topography, the corneal thickness was studied in the center, PCT at 1.5; 2, 3, 4 and 5 mm from the center in four meridians, as well as ΔPCT (PCT 3 mm - PCT 1.5 mm), the curvature of the anterior and posterior surfaces of the cornea and the depth of the anterior chamber. Aberrometry was used to obtain refractometry data and the curvature of the anterior surface of the cornea. The influence of the studied parameters on ΔIOP was evaluated.

RESULTS

ΔIOP correlated with CRF (r= -0.652), CH (r= -0.873), central corneal thickness (r= -0.293), PCT at all distances except 5 mm (r= -0.297; -0.287; -0.302; -0.303), with the strong and weak meridians of the anterior surface of the cornea (r=0.328; r=0.315), with the strong and weak meridians of the posterior surface, as well as the average curvature of the posterior surface (r=0.307; r=0.332; r=0.328). After step-by-step selection of the above parameters for creating a linear regression model for ΔIOP calculation, CH, CRF and PCT1.5mm remained in the model. The model describes ΔIOP with high accuracy (R2=0.974).

CONCLUSION

Biomechanical parameters of the cornea are the leading factor of applanation tonometry error. Individual linear dimensions of the cornea (thickness, curvature) have a lesser effect.

Для цитирования: Antonov A A, Volzhanin A V, Klinicheva E A, Karpilova M A. [Influence of corneal curvature and peripheral thickness on tonometry readings].. Vestnik oftalmologii. 2024. DOI: 10.17116/oftalma202414002243.