A prospective, observational evaluation of a new supraglottic airway: the PAXpress.

Mondello E, Casati A, Italian PAXpress Group

Оценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIM: Little information is available on efficacy and safety of the new supraglottic device, PA(Xpress), in anesthetized patients. The aim of this prospective observational study was to evaluate the use of this new supraglottic device in clinical practice.

METHODS

Ninety-one patients (mean age: 45+/-11 years) undergoing extremity or minor peri-phery procedures (mean duration 53+/-18 min), with a supraglottic airway were prospectively studied in 10 hospitals, using a simple questionnaire with data concerning anthropometric variables, surgical procedure, and occurrence of untoward events during PA(Xpress) placement, anesthesia maintenance, or postoperative period.

RESULTS

First attempt placement was achieved in 74 patients (82.2%), while 9 patients (10%) required 2 attempts, and 7 patients (7.7%) required more than 2 attempts. In only 2 patients (2%) airway control was not achieved with the studied device. The mean time required to achieve successful placement was 49+/-18 s (range 4-300 s). Mechanical ventilation was effectively maintained in all studied patients, and no severe adverse events were reported during the procedure. Blood on the device was observed in 54% of cases, and this was associated with an incidence of sore throat of 26% in the recovery area and 13% after 6 hours from the end of surgery.

CONCLUSION

Although further comparative, randomized studies should be advocated to better evaluate the use of this new supraglottic device, this prospective, observational study demonstrated that PA(Xpress) provides safe and effective airway control during mechanical ventilation in up to 98% of cases.

Для цитирования: Mondello E, Casati A, Italian PAXpress Group. A prospective, observational evaluation of a new supraglottic airway: the PAXpress.. Minerva anestesiologica. 2003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.