Finite Element Analysis of Stress Distributions for Mandibular Dental Implant-Supported Overdentures with Magnetic Attachments in Osteoporotic and Normal Bone.
Аннотация
PURPOSE: To compare the biomechanical responses of a normal mandible to an osteoporotic mandible with two-implant-supported magnetic attachments.
MATERIALS AND
METHODS
A 3D finite-element model of a two-implant-supported mandibular overdenture with magnetic attachments was developed, and normal and osteoporotic bone samples were prepared. Four types of load were applied to the overdenture in each model: 100 N vertical and oblique loads on the right first molar, and a 100 N vertical load on the right canine and incisors. Biomechanical behaviors of the peri-implant bone, implant, and mucosa were recorded. Maximum equivalent stresses and elastic strains were analyzed.
RESULTS
Equivalent elastic strain in osteoporotic cortical and cancellous bone was 9% to 71% and was 142% and 207% greater than in normal cortical bone, respectively. Equivalent elastic strain in the first molar oblique loading condition was 101% to 190% greater than in the first molar vertical loading condition.
CONCLUSIONS
Osteoporotic cancellous bone was weaker and less resistant to deformation than normal bone, and oblique loading was more harmful than vertical loading.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.