MedElevator Войти в исследования
← Поиск публикаций

DOI 10.22575/interventionalradiology.2024-0038

Медицинская библиотека · Interventional radiology (Higashimatsuyama-shi (Japan) · 2025

Evaluation of VIABAHNⓇ Endoprosthesis for Thoracic, Abdominal, and Pelvic Artery Injuries.

Anai Hiroshi, Tanaka Toshihiro, Nishiofuku Hideyuki, Ikeda Osamu, Abo Daisuke, Baba Yasutaka, Tajima Hiroyuki, Ota Toyohiro, Kichikawa Kimihiko

СтатьяПеревод в очереди

Аннотация

Purpose: To evaluate the efficacy and safety of the GOREⓇ VIABAHNⓇ Endoprosthesis (stent graft) for traumatic or iatrogenic vessel injury. Material and

Methods

This prospective, multicenter, cohort study for post-marketing clinical surveillance was conducted in Japan. Efficacy and safety endpoints included successful device implantation, primary and secondary hemostasis of injury, 1-month and 1-year survival rate, and at 1-, 6-, 12-months post-procedure, primary and secondary patency rates, antiplatelet drug administration, and serious adverse events.

Results

In 37 patients (mean age 73.3±12.5 years; 22 males; 59.5%), 38 vessel injuries were evaluated. Primary treated arteries were visceral, subclavian, and iliac arteries; 86.8% (33/38) were iatrogenic, and 13.2% (5/38) were traumatic vessel injury. Iatrogenic injury included pseudoaneurysm (63.6%), perforation (24.4%), rupture (15.2%), dissection (12.1%), or fistula (3.0%). Technical success in 36 patients (36/37; 97.3%) with 37 vessels (37/38; 97.4%), and primary and secondary hemostasis in all patients were achieved. Primary patency at 1-, 6-, and 12-months was 89.1%, 84.8%, and 80.6%, and secondary patency was 91.5%, 87.3%, and 82.4%, respectively. The Kaplan-Meier survival estimate was 94.4% through 1 month and 75.9% through 1 year. No device-related deaths occurred. Through 12 months, of the 19 serious adverse events (13 patients), three serious adverse events were judged by the investigator to be device-related (vascular stent graft stenosis) or procedure-related (device occlusion and cerebellar infarction).

Conclusions

The use of the VIABAHN stent graft for vessel injury was safe and effective in this study. Successful hemostasis was achieved through 1 month. Through 1 year, no device-related deaths were reported.

Для цитирования: Anai Hiroshi, Tanaka Toshihiro, Nishiofuku Hideyuki, Ikeda Osamu, Abo Daisuke, Baba Yasutaka, Tajima Hiroyuki, Ota Toyohiro, Kichikawa Kimihiko. Evaluation of VIABAHNⓇ Endoprosthesis for Thoracic, Abdominal, and Pelvic Artery Injuries.. Interventional radiology (Higashimatsuyama-shi (Japan). 2025. DOI: 10.22575/interventionalradiology.2024-0038.