Patient dose and associated exposure parameters in pelvic x-ray examinations: dependence on radiographic system.

Welarathna Sachith, Velautham Sivakumar, Sarasanandarajah Sivananthan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Technological differences between computed radiography (CR) and digital radiography (DR) systems can influence patient doses and exposure parameters in pelvic x-ray examinations. The presence of radiosensitive organs in the pelvic region underscores the need to optimize these parameters for both CR and DR systems. This prospective study aimed to compare the patient doses and exposure parameters for adult patients undergoing pelvic x-ray examinations using CR and DR systems, based on data from Sri Lanka. The study included data from 56 x-ray examinations, with 25 using CR and 31 using DR. Patient demographic characteristics and exposure parameters (kVp: kilovoltage peak, mAs: tube current-exposure time product) were recorded, and patient doses were measured in terms of the kerma-area product (PKA) using a PKA meter. Despite similar mean weight and body mass index (BMI), the CR systems showed significantly higher mean kVp (7.4%), mAs (16.4%), and PKA (29.7%) than the DR systems (CR - kVp: 73.2, mAs: 37.8, PKA: 2.29 Gy cm2; DR - kVp: 67.8, mAs: 31.6, PKA: 1.61 Gy cm2). The Mann-Whitney U test revealed statistically significant differences in PKA and kVp between the CR and DR systems (p < 0.05). Furthermore, even with lower patient weight and BMI, the mean mAs and PKA in this study were substantially higher than those reported in the literature for both CR and DR systems. These results suggest the need to optimize current mAs settings for the studied hospitals and introduce radiographic system-specific exposure parameters and reference dose levels for pelvic x-ray examinations in order to enhance patient protection.

Для цитирования: Welarathna Sachith, Velautham Sivakumar, Sarasanandarajah Sivananthan. Patient dose and associated exposure parameters in pelvic x-ray examinations: dependence on radiographic system.. Radiation and environmental biophysics. 2024. DOI: 10.1007/s00411-024-01080-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.