MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejro.2024.100578

Медицинская библиотека · European journal of radiology open · 2024

Reduced dose helical CT scout imaging on next generation wide volume CT system decreases scan length and overall radiation exposure.

Golbus Alexa E, Schuzer John L, Steveson Chloe, Rollison Shirley F, Matthews James, Henry-Ellis Joseph, Razeto Marco, Chen Marcus Y

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Traditional CT acquisition planning is based on scout projection images from planar anterior-posterior and lateral projections where the radiographer estimates organ locations. Alternatively, a new scout method utilizing ultra-low dose helical CT (3D Landmark Scan) offers cross-sectional imaging to identify anatomic structures in conjunction with artificial intelligence based Anatomic Landmark Detection (ALD) for automatic CT acquisition planning. The purpose of this study is to quantify changes in scan length and radiation dose of CT examinations planned using 3D Landmark Scan and ALD and performed on next generation wide volume CT versus examinations planned using traditional scout methods. We additionally aim to quantify changes in radiation dose reduction of scans planned with 3D Landmark Scan and performed on next generation wide volume CT.

METHODS

Single-center retrospective analysis of consecutive patients with prior CT scan of the same organ who underwent clinical CT using 3D Landmark Scan and automatic scan planning. Acquisition length and dose-length-product (DLP) were collected. Data was analyzed by paired t-tests.

RESULTS

104 total CT examinations (48.1 % chest, 15.4 % abdomen, 36.5 % chest/abdomen/pelvis) on 61 individual consecutive patients at a single center were retrospectively analyzed. 79.8 % of scans using 3D Landmark Scan had reduction in acquisition length compared to the respective prior acquisition. Median acquisition length using 3D Landmark Scan was 26.7 mm shorter than that using traditional scout methods (p < 0.001) with a 23.3 % median total radiation dose reduction (245.6 (IQR 150.0-400.8) mGy cm vs 320.3 (IQR 184.1-547.9) mGy cm). CT dose index similarly was overall decreased for scans planned with 3D Landmark and ALD and performed on next generation CT versus traditional methods (4.85 (IQR 3.8-7) mGy vs. 6.70 (IQR 4.43-9.18) mGy, respectively, p < 0.001).

CONCLUSION

Scout imaging using reduced dose 3D Landmark Scan images and Anatomic Landmark Detection reduces acquisition range in chest, abdomen, and chest/abdomen/pelvis CT scans. This technology, in combination with next generation wide volume CT reduces total radiation dose.

Для цитирования: Golbus Alexa E, Schuzer John L, Steveson Chloe, Rollison Shirley F, Matthews James, Henry-Ellis Joseph, Razeto Marco, Chen Marcus Y. Reduced dose helical CT scout imaging on next generation wide volume CT system decreases scan length and overall radiation exposure.. European journal of radiology open. 2024. DOI: 10.1016/j.ejro.2024.100578.