DOI 10.2174/0113892002274982231211102127
Reassessment of Radiation Exposure From Bone Scintigraphy.
Аннотация
AIM: This study was aimed to re-determine the radiation dose rate emitted from the patients who underwent bone scintigraphy.
MATERIAL AND
METHODS
A mean of 20.87±2.54 mCi 99mTc-MDP was injected into patients. A GM counter was used to measure dose rates in 3 different periods, at intervals of 25, 50, 100, 150, and 200 cm from the patient's anterior for head, thorax, abdomen, and pelvis levels. Measurements were used to determine patient-induced environmental doses and radiation doses to personnel/patient relatives.
RESULTS AND DISCUSSION: There were strong correlations between mean dose rate (mRh-1mCi-1) and time at all regions and distances. The received dose for staff was calculated between a range of 0.01-0.02 mSv/mCi per patient. The total dose to be received by the companion was estimated to be between 0.019-0.039 and 0.011-0.022 mSv for public and personal vehicle transportation, respectively. The radiation dose exposed by nurses (4th, 6th, and 8th hours after injection) was found to be 0.012-0.064, 0.006-0.038, and 0.002-0.018 mSv/- patient, respectively.
CONCLUSION
The fact that the doses of personnel and patient relatives in the study were below the legal limits shows that the study was carried out within a safe range. However, in terms of radiation protection, it is necessary to limit the time spent with the patient as much as possible and increase the distance. Since the dangers of low radiation dosages are unknown, there is a need to inform the patient's relatives and staff about the potential risks.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.