DOI 10.14309/ctg.0000000000000976
Real-Time Characterization of Colonic Polyps: A Multicenter Prospective Study Evaluating the CAD-EYE System in Screening Colonoscopies.
Аннотация
INTRODUCTION
Polypectomy-related costs could potentially be reduced through optical diagnosis strategies, such as "diagnose-and-leave" and "resect-and-discard." Artificial intelligence, using computer-aided diagnosis (CAD), may provide a reproducible optical diagnosis of colorectal lesions. This study aimed to assess the performance of the CAD-EYE system in the real-time characterization of colonic polyps.
METHODS
We conducted a cross-sectional, multicenter study evaluating the CAD-EYE system in patients undergoing screening colonoscopies at 5 French centers. CAD-EYE predictions and assessments by endoscopists (hyperplastic vs neoplastic) were compared with histopathology results. The primary outcome was the sensitivity of CAD-EYE for predicting neoplastic polyps, compared with the predefined threshold of 85%. The secondary outcomes were the specificity, positive predictive value, negative predictive value, endoscopists' performance, and polyp detection rates.
RESULTS
Of 398 polyps analyzed, 343 were included in the primary analysis. CAD-EYE characterization was feasible in 96% of cases. The sensitivity was 0.80 (95% confidence interval, 0.74-0.85), which failed to achieve the predefined threshold of 85% ( P = 0.064). The specificity, negative predictive value, and positive predictive value were 0.79, 0.64, and 0.90, respectively. Performance was higher for diminutive rectosigmoid polyps (DRSPs). Endoscopists showed higher sensitivity than CAD-EYE (0.90 vs 0.80, P = 0.001). CAD-EYE-assisted colonoscopies detected more polyps per procedure (3.3 vs 2.3, P < 0.001) than endoscopists alone.
DISCUSSION: The performance of CAD-EYE was insufficient for the characterization of neoplastic colonic polyps. CAD-EYE performed better for DRSPs. AI seems to be beneficial for polyp detection.