Quantification of the Proton Density Fat Fraction and Iron Content: A Comparative Study Between 3.0 T and 5.0 T MRI.
Аннотация
BACKGROUND
Proton density fat fraction (PDFF) and R2* are noninvasive MRI biomarkers for quantifying fat and iron in abdominal organs. While 3.0 T MRI is widely used clinically, 5.0 T may offer improved accuracy and reliability. With the increasing availability of 5.0 T, assessing its feasibility and utility for quantifying PDFF and R2* in abdominal organs is needed.
PURPOSE: To determine the agreement of the PDFF and R2* in the liver, pancreas, kidney, and paraspinal muscle at 2 different field strengths.
STUDY TYPE: Prospective.
POPULATION: A total of 127 participants, including 60 healthy volunteers and 67 with metabolic dysfunction-associated steatotic liver disease (100 women and 27 men, 52 ± 9 years old).
FIELD STRENGTH/SEQUENCE: 3.0 T and 5.0 T; Chemical shift-encoded multi-echo gradient echo sequence.
ASSESSMENT: PDFF and R2* values in the liver, pancreas, kidney, and paraspinal muscle were measured by three observers to evaluate the interobserver and interfield agreement using two 3.0 T scanners (scanner 1 and scanner 2) and one 5.0 T scanner.
STATISTICAL TESTS: Linear regression, Bland-Altman analyses, and intraclass correlation coefficients (ICCs). The p-value < 0.05 indicated statistically significant.
RESULTS
PDFF and R2* at 5.0 T were strongly correlated with those at two 3.0 T scanners for all organs (R 2 = 0.905-0.998 for PDFF; 0.831-0.991 for R2*), indicating high interfield ICCs (0.892-0.993 for PDFF; 0.910-0.981 for R2*). Comparisons between 5.0 T and two 3.0 T scanners showed good interfield agreement for PDFF (mean bias, -0.57% to 0.03%) while a higher bias for R2* (mean bias, -16.53 to -2.64 s-1, 95% LoA, -34.28 to 1.21 s-1) at 5.0 T compared to the comparison between 3T scanners.
DATA
CONCLUSION
5.0 T revealed high interfield agreement between the PDFF and R2* with 3.0 T, which could provide reliable quantification of fat and iron contents in abdominal organs.
EVIDENCE LEVEL: 2.
TECHNICAL EFFICACY: Stage 2.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.