DOI 10.1111/aogs.12649

Медицинская библиотека · Acta obstetricia et gynecologica Scandinavica · 2015

Transperineal ultrasonography for labor management: accuracy and reliability.

Yuce Tuncay, Kalafat Erkan, Koc Acar

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare ultrasound measurements and clinical assessments of cervical dilatation, fetal head station and fetal head position.

DESIGN

Prospective, observational study.

SETTING

Tertiary care center (Ankara University Hospital), labor ward.

POPULATION: Forty-three women in labor.

METHODS

Women were prospectively evaluated through simultaneous examinations (79 in total). Dilatation, head station, and position were recorded via digital examination whenever possible and were compared with ultrasonographic measurements of dilatation, angle of progression, head-perineum distance, and head position by two independent operators. Operators were blinded to each other's findings.

MAIN OUTCOME MEASURES: Agreement between sonographic measurements and digital examinations.

RESULTS

Ultrasonographic assessment of dilatation correlated significantly with digital examination. The interclass correlation coefficient of the two methods was 0.82 (95% confidence interval 0.73-0.88). Ultrasonography showed a mean lower cervical dilatation of 10 mm (95% limits of agreement: -36 to 16 mm) compared with digital examination. The angle of progression and head-perineum distances correlated moderately with the palpated head station (Pearson's correlation coefficients: 0.55 and -0.42, respectively; p < 0.001). Position could only be assessed in 40 of 79 examinations by digital examinations and the results showed low agreement with ultrasound findings.

CONCLUSION

We observed good agreement between clinical and ultrasound assessment of dilatation, moderate agreement in assessing fetal station and low agreement assessing head position.

Для цитирования: Yuce Tuncay, Kalafat Erkan, Koc Acar. Transperineal ultrasonography for labor management: accuracy and reliability.. Acta obstetricia et gynecologica Scandinavica. 2015. DOI: 10.1111/aogs.12649.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.